23
Figure 12. Objective Band Heater Selection Chart.
If your system is also equipped with the Objective Heater, (OBJ-COLLAR ####,
optional) its objective collar should be installed before connecting the heater to rear
panel of the Temperature Control Unit. Follow the steps shown in Figure 13
Remove the microscope objective and place it on the desk
Place the orange colored Heating Band around the microscope Objective, cover
it with the Velcro tape and fix the Velcro Tape as illustrated in Figure 13
Then install the Objective to the microscope nosepiece, rotate the Objective
carefully avoiding damaging the wires or the flexible heating band.
Avoid holding the Objective Heater Band by lead wires.
Do not modify or cut in any way the Objective Heating Band in the attempt to fit
objective of different size.
Make sure that the Objective Heating Band is wrapped around the objective before
turning it on. If the Objective Heating Band is on while not wrapped around an
objective, it can overheat and can be damaged.