39
Once you have entered the Main Menu pressing the ‘Menu’ Key:
Move the cursor to the item ‘Offset’ with the ‘Arrow’ Keys and press ‘Enter’.
Here you can read the offset values. If you are using a CUSTOM PLATE ADAPTER, not included
in the list of standard adapters or if the room temperature differs more than 2°C from 23°C
(temperature of the factory calibration) follow this procedure: To change the ‘Reading Offset’
select ‘Custom’ configuration. Here you can change the Reading Offset of the base plate, lid
and humidity module temperatures in order to have the sample at the set point temperature.
Select this configuration only when using a custom plate adapter, or if the room temperature
differs more than 2°C (See paragraph 13.5.2)
Tip
►
Keep in mind that thermal control settings change when changing the adapter in use.
Hence, every time you change adapter you must input the correct base offset using the
procedure of Figure 24 and the offset values reported in Table 1. In case you purchase
additional incubating chamber models make sure to input the correct base offset prior to your
experiment.
Menu 1/3
à
Offset Mode
Offset
Objective Heater
8
Offset Mode
à
Mode Standard
Ref @ 37°C: -1.3
Offset Mode
Mode Standard
à
Ref @ 37°C:
-1.3
â
8
8
Enter to
confirm
Move the cursor to
currect base offset
-
Exit
Figure 24. How to change the base offset value.
Press ‘Menu’ Key
Press ‘Enter’ with the cursor on the item ‘Offset Mode’
Move the cursor to the item ‘Ref @ 37°C’ and press ‘Enter’
Move the cursor to the correct base offset while it is flashing and press Enter to
confirm the chosen offset value.
Press ‘Menu’ Key to exit from the Sub Menu. Press ‘Menu’ key to go to the ‘Main
Page’.