40
Please note that the only offset value to change is the base offset. Lid and HM
offsets will be calculated as a function of the base offset and you will see them in
‘Offset’ submenu. (See Figure 23).
13.5.2
Custom Offset Mode
In order to select the Custom Offset Mode, see Figure 25. If the room temperature differs
more than 2°C from 23°C (temperature of the factory calibration) or if you use a custom plate
adapter, non-included in the list reported in Table 1, Okolab recommends this working mode
recalibrating the chamber following the procedure reported in paragraph 12.
Menu 1/3
à
Offset Mode
Offset
Objective Heater
8
Offset Mode
à
Mode
Custom
Ref @ 37°C: -1.3
Offset Mode
Mode Custom
8
Enter to
confirm
-
Exit
8
Figure 25. How to choose the Custom Offset Mode
From the main menu press the ‘Menu’ key on the front panel.
Press ‘Enter’ with the cursor on ‘Offset mode’
Press ‘Enter’ with the cursor on ‘Mode’
Press ‘Up/down’ until you see ‘Custom’
Press ‘Enter’ to select ‘Custom’ as the configuration in use.
Press ‘Menu’ Key to exit from the Sub Menu.
Press ‘Menu’ key to go to the ‘Main Page’.
Once you have selected ‘Custom’ mode, you can change the base, lid and HM offset iterating
the same procedure shown in Figure 26 for the three devices.