30
Once you have selected Standard mode, you can change the set point according to your
needs:
From the ‘Main Page’, press ‘Enter’ to change the Set point Temperature.
Press ‘up’ or ‘down’ arrows button to increase or to reduce the Set Point
Temperature while it is flashing.
Press ‘Enter’ to confirm the new Set Point Temperature.
Tip
►
Reading Offsets were evaluated at ambient temperature of 23+/-1°C for each Set Point
temperature within the working range 25÷50°C and stored in the EEPROM memory of the
Control Unit. This enables the user to change the Set Point temperature within the working
range 25÷50°C. It also allows maintaining the sample temperature at the Set Point
Temperature without the need of any additional calibration.
The Reading Offset values vary with the Set Point temperature and with the
chamber and sample adapter in use.
Advantages of this approach:
The incubating system is Easy to use
Fast experiment start up
Well suited for multi users
Sample temperature accuracy of +/-0.3°C from the Set Point Temperature.
Disadvantage:
It needs of calibration in case of the ambient temperature is different from 23°C
+/-1°C.
The default values are optimized for T
Ambient
23.0 ± 1.0°C and active Humidifying Module
heater.
Offset values are evaluated according to the following:
1.
Maintaining sample temperature error within ± 0.3 °C
2.
Avoiding condensation on the lid glass
3.
Improving the gas stream humidity level