Appendix
A-6 A-6
19. Attach the controller to the base platform by tightening the two mounting bolts at
the bottom of the controller housing. Be careful not to over tighten the mounting
bolts, since this may damage the power switch. Route all cables through the cut-out
in the back of the controller so that they are not pinched between the incubator and
the controller housing.
20. Insert the servo-controlled humidifier fill tray and lid assembly.
21. Re-mount the incubator to the table top using the four mounting knobs removed
earlier.
22. Connect the humidifier power cord and secure it to rear left corner of the base with
the mounting screw and strain relief clip provided (see Figure A-4).
23. Re-attach the tilt knob assembly, if applicable.
24. Perform the check-out procedure in section 3 and the electrical safety test in
section 6.
Figure A-4
Power cord attachment
Strain relief clip
Humidifier
controller
receptacle
Power cord
Mounting screw
CI.18.004