Appendix
A-2 A-2
m
IEC 601-1 Type B Isolation
Mode of Operation: Designed for continuous operation.
Protection against
ingress of water:
Not protected.
Leakage Current:
< 100µA for 120V models
< 500µA for 230V models
Self Test:
The microprocessor performs self tests and software verification functions
when the power is first switched on.
Environmental
Operating Temperature Range: 20°C to 30°C
Transport & Storage Range: -25°C to 60°C
Transport, Operating & Storage Pressure Range: 500 hPa to 1060 hPa
Operating Relative Humidity Range: 10 to 60%
Transport and Storage Relative Humidity Range: 0 to 95%
Electromagnetic Compatibility (EMC)
EC 801-2
Electrostatic Discharge Requirements Level 2
IEC 801-3
Immunity to Radiated Radio-frequency Electromagnetic Fields Level 2
IEC 801-4
Electrical Fast Transient Burst Requirements Level 2
IEC 801-5
Surge Immunity Requirements Level 2
IEC 801-6
Immunity to Conducted Disturbances Induced to Radio-frequency Fields
above 9 kHz Level 2
CISPR 11
Limits and Methods of Measurements of Radio Interference Characteristics of
Industrial, Scientific and Medical Radio-frequency Equipment Class A
Mechanical
Tray Capacity: 1000 ml
Summary of Contents for Ohio Care Plus
Page 20: ...2 General Information 2 4 2 4 Notes...
Page 24: ...4 Using the Servo controlled Humidifier 4 2 4 1 Notes...
Page 46: ...6 Service 6 18 6 18...
Page 53: ...8 Schematics 8 1 Figure 8 1 Wiring Diagram Servo controlled Humidifier 8 Schematics CI 18 059...
Page 55: ...8 Schematics 8 3 Figure 8 3 Digital section of the Control Display Board Assembly CI 18 061...
Page 56: ...8 Schematics 8 4 Figure 8 4 Analog section of the Control Display Board Assembly CI 18 062...
Page 58: ...8 Schematics 8 6 Figure 8 6 Display section of the Control Display Board Assembly CI 18 064...
Page 67: ......