Effective April 2006
41
SCREEN 5
This Meter; REGULATED AIR
involves reading your incoming
Regulated Air from the compressor.
Measurements are in psig, there is
both a analog reading and a digital
gauge. The digital reading is exact,
while the analog meter acts as a real
meter. If the dial falls out of the
green area an error will appear. The
normal operating variance is between
40 and 75 psi.
This Meter; WASTE GAS measures
the psig of your Waste Gas. The
same details hold true for this gauge
as in the REGULATED AIR gauge
except for the normal operating
variance that is between 0 and 20 psi.
This Meter; OXYGEN PRESSURE
measures the Oxygen Pressure (psig)
in your back pressure regulator. The
same details hold true for this gauge
as in the REGULATED AIR gauge
except for the normal operating
variance which is 45 – 62 psi.
This Meter; OXYGEN PURITY
measures the Oxygen Purity that the
generator is producing. The same
details hold true for this gauge as in
the REGULATED AIR gauge except
that the Oxygen being generated
should be between 85 – 98%.
If you want to go back to the
Previous Screen/Main Screen all you
have to do is touch the Blue Area.
If you want to advance to the Next
Screen/Pressure Screen all you have
to do is touch the Blue Area.
Summary of Contents for OG-75
Page 15: ...Effective April 2006 12 Sieve Re packing Diagram...
Page 26: ...Effective April 2006 23 Process Flow Schematic...
Page 27: ...Effective April 2006 24 Process Flow Schematic Touchscreen Option...
Page 29: ...Effective April 2006 26 External Components Drawing 22 G G 22...
Page 33: ...Effective April 2006 30 Internal Components Drawing SMC SMC Direct 05 LOGIC Koyo SMC...