SCREEN 7
This particular alarm is triggered
when the Regulated Air is below
40 psi.
NOTE! These are triggered alarms that will only flash when the machine is cycling.
Initially when the generator begins cycling these errors will occur.
This particular alarm is triggered
when the Regulated Air is above
80 psi.
This particular alarm is triggered
when the Oxygen Pressure is
below 40 psi.
This particular alarm is triggered
when the Oxygen Pressure is
above 65 psi.
This particular alarm is triggered
when the Oxygen Purity is below
85%.
This particular alarm is triggered
when the Oxygen Purity is above
96%.
This particular alarm is triggered
when the dispersed gas from the
muffler is out of the range of
0 – 20 psi
This button allows you to
look at what caused the errors
in the system. This is a
database of errors.