Effective April 2006
34
The next screen is the Gauge Screen, which shows all the gauges of operation.
The Regulated Air guage is in psig, the normal variance of this gauge is between 40
(low) and 80 (high) psig. The Waste Gas guage is in psig, the normal variance of this
gauge is between 0 (low) and 20 (high) psig. The Oxygen Pressure guage is in psig,
the normal variance of this gauge is between 45 (low) and 65 (high) psig. Oxygen
Purity is in percent (%), the normal variance of this gauge is between 85 (low) and 98
(high) psig.
The Pressure Screen is the next screen, this screen contains a graph of the 4
gauges over time. Every five minutes a reading is taken. The legend on the bottom of
the screen contains the coordinating colors of the meters..
The Alarm Screen contains 7 triggered alarms that only occur when the machine
is cycling. These must be ignored when the generator is initially started up because the
machine must warm up. The same principle holds true for the alarms on the bottom of
the screen. The button that says Touch To View Alarm History, displays the database
of the warnings on the bottom of the screen, which tells you to go to the alarm screen,
and the triggered errors will be displayed if the machine is cycling.
This Final Screen contains all the general information of the generator, and how
to contact OGSI.
Summary of Contents for OG-75
Page 15: ...Effective April 2006 12 Sieve Re packing Diagram...
Page 26: ...Effective April 2006 23 Process Flow Schematic...
Page 27: ...Effective April 2006 24 Process Flow Schematic Touchscreen Option...
Page 29: ...Effective April 2006 26 External Components Drawing 22 G G 22...
Page 33: ...Effective April 2006 30 Internal Components Drawing SMC SMC Direct 05 LOGIC Koyo SMC...