6. Cleaning, maintaining, and repairing
6.2. Model ASC-1
1. Insert a flat-bladed screwdriver into the slot at the side panel.
1
ASC-1: View on the right side panel; (1) slot
2. Slide the side panel towards the device’s front.
3. First pull the side panel near the device’s rear and then near the device’s
front to the side.
4. Replace the air filter with an equivalent one. Alternatively blow the air
filter off.
5. Reassamble the air filter and the side panel in reverse order.
6.2.2. REPLACING THE BATTERY
The battery mainly supplies the internal clock. The internal clock provides date
and time for the LOG-files. If the internal clock is not running then you will not
be able to correlate events on the time scale. But this may be necessary for
troubleshooting.
TARGET AUDIENCE AND QUALIFICATION
Experts
HAVE READY
Battery; see table A.1 on page 93
Non-conductive tweezers
Flat-bladed screwdriver; width max. 5 mm