4. Preparing for use
4.4. Commissioning
REQUIREMENTS
The ASC is installed and connected to your devices as described in sec-
tion 4.3 on page 27.
The laser unit complies with the standards that are required by the dec-
laration of incorporation of each component.
PROCEDURE
1. Check with your network administrator if the factory settings can be used
within your Local Area Network or PC. The factory settings are:
IP-address
192.168.1.42
subnet mask
255.255.255.0
2. Check with your network administrator that exceptions are made at ports
80, 1610, 1611 and 1612 in all active firewall and security software to ensure
access to the ASC.
3. If the factory settings can be used then the ASC is connected to your
Local Area Network or PC now and the procedure ends here then.
– OR –
If the factory settings
can not
be used then
a) Request an IP-address and a subnet mask for the ASC from your
network administrator and note them down here.
IP-address
|
_
|
_
|
_
|
.
|
_
|
_
|
_
|
.
|
_
|
_
|
_
|
.
|
_
|
_
|
_
|
Subnet mask
|
_
|
_
|
_
|
.
|
_
|
_
|
_
|
.
|
_
|
_
|
_
|
.
|
_
|
_
|
_
|
b) At the ASC front panel, turn the key to horizontal position.
c) Press (hard) the standby/on push-button (ARGES head).
The blue light starts an upward motion indicating that the system is
starting. Once the system is startup sequence is completed the push-
button is illuminated steadily and the right display indicates that the
system is ready.