Getting
started
XT V 160 X-ray System
XTM0011-F2
37
There is no recommended time to hold the X-ray source at the maximum kV during
auto-conditioning. The longer it is held there, the more stable the X-ray source will be at
higher kV values.
If auto-conditioning fails repeatedly, then it generally means that the
X-ray source needs cleaning.
5.3
Check the status of the system
The Inspect-X
System Status
window shows the detailed status of the system and whether
interlocks have been made to allow the manipulator to be moved within the cabinet and
X-rays to be generated. If interlocks are not made, and the doors on the inspection system
are closed correctly, the information in the
System Status
window may help to identify
possible problems.
The System Status also appears at the top of the Inspect-X window:
To view the status of the system, follow these steps:
1. Click
.
The
System Status
window opens.
2. If the X-ray controller interlock is not made, check that:
•
The INTERLOCK key is in the ON position.
•
All cabinet doors are closed.
•
The X-ray ON Lamp is operating correctly.
•
The turbo pump is up to speed (shown in the
System Status
window).
•
The vacuum is below the set point (shown in the
System Status
window).
•
The coolant flow and level is correct.
3. If the manipulator interlock is not made, check that:
•
The INTERLOCK key is in the ON position.
•
All cabinet doors are closed.
4. Click the
Close
icon (X) to close the
System Status
window.
5.4
Inspect a sample
This section provides an overview of the main steps required to inspect a sample and
capture and save an image. It assumes the system is switched on and the source is
conditioned and ready for use.
For more detailed instructions on using Inspect-X to inspect a sample,
refer to the
Inspect-X Online Help
or
User Manuals
.
5.4.1
Home the manipulator
Summary of Contents for XT V 160
Page 1: ...X Tek X ray and CT Inspection XT V 160 X ray System Operator Manual XTM0011 F2 ...
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ......
Page 22: ......
Page 26: ......
Page 32: ......
Page 83: ...Maintenance XT V 160 X ray System XTM0011 F2 73 ...
Page 90: ...Troubleshooting 80 XTM0011 F2 XT V 160 X ray System ...
Page 93: ......