Warnings
XT V 160 X-ray System
XTM0011-F2
21
•
The backing vacuum pump in this system uses mineral oil. For a brief period following
switch on, and in the event of a malfunction, oil mist may be produced. Ensure all
suitable precautions are taken when handling and refer to the relevant Material Safety
Data Sheet.
•
The X-ray cabinet uses lead for shielding. The lead is in solid form and all exposed
surfaces are covered or painted so an operator should not come into direct contact.
However, ensure all suitable precautions are taken and refer to the relevant Material
Safety Data Sheet (MSDS).
•
Some X-ray source windows are made from beryllium. Care should be taken to avoid
scratching or abrading the window when removing as this may release toxic dust.
•
Coolant oil is not classified as flammable but will burn.
•
Self-sealing connectors on coolant hoses may leak a small quantity of coolant when
disconnected. This should be cleaned up using suitable protective equipment.
The following warnings are identified as mandatory actions which must
be followed to prevent a hazard:
•
Fluid products must not be allowed to get into the eyes and prolonged or repeated
contact with the skin must be avoided. Splashes into the eyes or onto the skin must
be washed away immediately with plenty of clean water, and medical advice should
be sought immediately.
•
Chemicals must be kept away from food and drink.
•
Inflammable products such as isopropyl alcohol (IPA, propanol) must only be used in
a well-ventilated area and away from any source of ignition.
•
Cooling systems contain Hexid A4 heat transfer fluid or mineral oil (Shell Diala). While
the coolant itself is not considered hazardous, best practice dictates the use of gloves
and goggles while handling. Use only pre-mixed Hexid A4 coolant to minimise
biological growth and prevent corrosion. Ensure all suitable precautions are taken
when handling and refer to the relevant Material Safety Data Sheet.
•
When working with coolant or mineral oil clean up any spills to prevent a slipping
hazard. Fluids should only be replaced by trained personnel. Comply with local
regulations when disposing of used fluids.
•
Beryllium is highly toxic in particulate form; always wear gloves when touching the
target window assembly.
•
Always observe local health and safety regulations relating to the handling and
disposal of beryllium and refer to the relevant Material Safety Data Sheet.
The following warnings are identified as prohibited actions that must
not be performed:
•
Any fumes produced by the system in the event of a failure, or by items in the spares
kit, must never be inhaled.
•
Do not scratch the target window or try to clean it with an abrasive material since this
may produce beryllium dust.
•
Do not allow beryllium dust to enter the body through cuts, inhalation or ingestion.
•
Never attempt to modify the beryllium window.
Summary of Contents for XT V 160
Page 1: ...X Tek X ray and CT Inspection XT V 160 X ray System Operator Manual XTM0011 F2 ...
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ......
Page 22: ......
Page 26: ......
Page 32: ......
Page 83: ...Maintenance XT V 160 X ray System XTM0011 F2 73 ...
Page 90: ...Troubleshooting 80 XTM0011 F2 XT V 160 X ray System ...
Page 93: ......