6 Maintenance
6.1
Preventive maintenance
Checks should be carried out at frequent intervals, dependent on the duty cycle of the
system and the type of environment. Please consult with your company health and safety
officer, or local authority to establish a suitable routine inspection procedure. The following
should be regarded as a guide to a minimum standard.
If you have any queries about maintenance procedures please contact Nikon Metrology.
Routine maintenance is only to be carried out by trained personnel.
Servicing is only to be performed by trained service engineers.
6.1.1
Interlock testing
The preventive maintenance schedule should include this interlock testing procedure:
Equipment Interlock testing
Name
Location
Activation
method
Verification
method
Test
frequency
Comment
Pass/Fail
Door
Upper and
lower door
corner
Open the
door
Try to move
the
manipulator.
Should not
respond to
commands.
Weekly
Door
Upper and
lower door
corner
Open the
door
Ensure the
interlock
status is
reported to
inspect-X
properly for
the X-ray
source
interlocks as
well as the
manipulator
Weekly