Getting
started
40
XTM0011-F2
XT V 160 X-ray System
Once the sample is loaded and you have set the initial X-ray conditions, you can switch
X-rays on. When you do this, there is a delay before the X-rays are generated by the X-ray
source in the inspection system.
This is the pre-warning delay and is a safety feature. The length of the pre-warning delay is
set to two seconds by default.
By default the pre-warning delay only occurs if the system door has
been opened since the last X-ray On.
To switch X-rays on, follow these steps:
1. In the X-ray zone, click
Switch X-rays On
.
The status indicator changes to indicate that X-rays are on:
2. While X-ray are on, conditions can be altered, as required:
•
The higher the kV is set, the greater the penetration.
•
The higher µA is set, the greater the brightness.
5.4.5
Display a live image
Switches the live imaging mode on
Switches the live imaging mode off
You can display a live image of an object under inspection on only one of the
Image
tabs in
Inspect-X. Use the other tabs to display captured images or images that you have saved
previously and re-loaded.
The live image displays a green border to distinguish it from captured
or loaded images. The border is red when X-rays are switched on.
To view a live X-ray image, follow these steps:
1. Ensure X-rays are switched on.
2. In the image window, click an
Image
tab. The live image displays on the tab you select.
3. On the image toolbar, click
.
The X-ray image on the selected tab is the view of the object, providing the manipulator
has been moved to view the object. If you cannot see an image, switch the X-rays on.
Live image options for area detectors
When you are viewing a live image, you can use the
C.Clear
frame averaging and
image-processing functions to set the quality of the image. The
C.Clear
imaging profile that
Summary of Contents for XT V 160
Page 1: ...X Tek X ray and CT Inspection XT V 160 X ray System Operator Manual XTM0011 F2 ...
Page 2: ......
Page 4: ......
Page 6: ......
Page 8: ......
Page 22: ......
Page 26: ......
Page 32: ......
Page 83: ...Maintenance XT V 160 X ray System XTM0011 F2 73 ...
Page 90: ...Troubleshooting 80 XTM0011 F2 XT V 160 X ray System ...
Page 93: ......