7 Registering Data for Optical Members
- 52 -
(3) Filter Block
You can register optional data for up to nine filter blocks.
If an attached filter block is not listed in the presets, register the filter block.
1
Press the MODE key, and the SETTING MENU screen will
appear on the LCD.
Select “Filter” using the
S
key or the
T
key and then
press the [
] key.
SETTING MENU
Objective
Condenser
Filter
Exciter
Barrier
Combination(Obj.>Condenser)
Combination(Filter>EX/BA)
More
2
Select “Edit” using the
S
key or the
T
key and then press
the [
] key.
FILTER
Name
1[UV-1A ] EX : EX365/10
2[V-1A ] DM : DM400
3[BV-1A ] BA : BA400
4[B-1A ]
5[G-1A ]
6[DIA ]
Edit
3
Select one location out of nine locations to register the
optional data using the PREV or NEXT key.
EDIT FILTER DATA
Location:
[
Other1]
Name :
[ ]
EX :
[――――――――――] Edit
DM :
[――――――――――] Edit
BA :
[――――――――――] Edit
4
Move the cursor to “Name” using the
S
key or the
T
key.
Type in the name of the filter block using the
numerical/alphabetical keys and the
W
key or the
X
key.
You can type in up to 10 characters.
EDIT FILTER DATA
Location:
[Other1]
Name :
[ABC
]
EX :
[――――――――――] Edit
DM :
[――――――――――] Edit
BA :
[――――――――――] Edit
5
Move the cursor to “EX” using the
S
key or the
T
key.
Select an excitation filter of the filter block using the PREV
or NEXT key.
If the excitation filter is not listed in the presets, register the
optional data for the excitation filter.
To register the optional data, move to “Edit” using the
W
key or the
X
key and then press the [
] key.
For details, see steps 3 forward for V-7 (5) Excitation
Filter.
EDIT FILTER DATA
Location:
[Other1]
Name :
[ABC ]
EX :
[
EX365/10 ] Edit
DM :
[――――――――――] Edit
BA :
[――――――――――] Edit
6
Move the cursor to “DM” using the
S
key or the
T
key.
Select a dichroic mirror for the filter block using the PREV
or NEXT key. If the dichroic mirror is not listed in the
presets, register the optional data for the dichroic mirror.
To register the optional data, move to “Edit” using the
W
key or the
X
key and then press the [
] key.
For details, see V-7 (4) Dichroic Mirror.
EDIT FILTER DATA
Location:
[Other1]
Name :
[ABC ]
EX :
[EX365/10 ] Edit
DM :
[
DM400 ] Edit
BA :
[――――――――――] Edit