VI Maintenance Menu
3 Testing Hardware of Remote Control Pad
- 61 -
VI
3
Testing Hardware of Remote Control Pad
The following tests are available to evaluate the hardware of the remote control pad.
Key input test
Checks to see if the input of each key is read correctly
LED ON/OFF test
Checks to see if all the LEDs are controlled correctly
LCD indication test
Checks to see if the LCD indication is controlled correctly and tests the quality of the
LCD
(1) Key Input Test
1
Turn on the power to the HUB controller while holding
down the MODE key.
After the MAINTENANCE MENU screen appears, release
the MODE key.
Select “TEST 2” using the
S
key or the
T
key and press
the [
] key.
MAINTENANCE MENU
1.Version
2.TEST1
3.
TEST2
4.PAD Data
5.HUB Data
6.Control Data
a.All Control Data
b.Revoiving Nosepiece
c.Condenser Cassette
d.Light Path
e.Filter Block
f.Exciter Filter
g.Barrier Filter
h.AUX Filter
2
Move “
→
” using the
S
key or the
T
key to select “KEY.”
Press the [
] key.
3
Press the keys (except the [
] key) one by one.
When the input for a key is read correctly, the key position
in the key layout on the screen will be indicated in black.
4
Press the [
] key to end the key input test.
5
Press the MODE key to return to OPERATION screen.