VII Troubleshooting
5 Voltage of dia-illumination lamp cannot be adjusted.
- 75 -
VII
5
Voltage of dia-illumination lamp cannot be adjusted.
Cause
Action
The dia-illumination lamp power supply is not
connected to the HUB controller correctly.
Turn OFF the power of the HUB controller. Disconnect the
dia-illumination lamp power supply from the HUB controller and connect it
again.
See “I-1 (1) Connecting Electrical Components to HUB Controller.”
EXTERNAL switch of the dia-illumination lamp
power supply is OFF.
Turn ON the EXTERNAL switch of the dia-illumination lamp power
supply.
See “I-1 (2) Turning ON EXTERNAL Switch of Dia-Illumination Lamp
Power Supply.”
The remote control operation is set to “OFF.”
Set the remote control operation to “ON.”
See “V-4 Setting Remote Control ON/OFF.”
The dia-illumination lamp is OFF.
Turn ON the dia-illumination lamp.
See “II-9 (1) Turning Lamp ON/OFF.”
6
Opening/closing status of Uniblitz shutter doesn’t coincide
with LED indication.
Cause
Action
The Uniblitz shutter controller is not in the N.C.
mode.
Place the Uniblitz shutter controller into N.C. mode.
See “I-1 (3) Placing Uniblitz Shutter Controller into N.C. Mode.”
7
Objective information is not indicated.
Cause
Action
The revolving nosepiece is not connected to the
HUB controller correctly.
Turn OFF the power to the HUB controller. Disconnect the revolving
nosepiece from the HUB controller and connect it again.
See “I-1 (1) Connecting Electrical Components to HUB Controller.”
Indication of information for the objective is OFF.
Turn ON the indication function for the information on the objective.
See “V-2 Setting Indication on Operation Screen.”
8
Condenser module information is not indicated.
Cause
Action
The condenser module changeover is not
connected to the HUB controller correctly.
Turn OFF the power to the HUB controller. Disconnect the condenser
module changeover from the HUB controller and connect it again.
See “I-1 (1) Connecting Electrical Components to HUB Controller.”
Indication of the information for the condenser
module is OFF.
Turn ON the indication function for the information on the condenser
module.
See “V-2 Setting Indication on Operation Screen.”
9
Filter block information is not indicated.
Cause
Action
The filter block changeover is not connected to the
HUB controller correctly.
Turn OFF the power to the HUB controller. Disconnect the filter block
changeover from the HUB controller and connect it again.
See “I-1 (1) Connecting Electrical Components to HUB Controller.”
Indication of the information for the filter block is
OFF.
Turn ON the indication function for the information for the filter block.
See “V-2 Setting Indication on Operation Screen.”