V Operation and Indication Functions
7 Registering Data for Optical Members
- 53 -
V
7
Move the cursor to “BA” using the
S
key or the
T
key.
Select a barrier filter for the filter block using the PREV or
NEXT key. If the barrier filter is not listed in the presets,
register the optional data for the barrier filter.
To register the optional data, move to “Edit” using the
W
key or the
X
key and then press the [
] key.
For details, see steps from 3 forward of V-7 (6) Barrier
Filter.
EDIT FILTER DATA
Location:
[Other1]
Name :
[ABC ]
EX :
[EX365/10 ] Edit
DM :
[DM400 ] Edit
BA :
[
BA400 ] Edit
8
To continue registration of optional data, move the cursor
to “Location” using the
S
key or the
T
key and repeat
steps 3 - 7 above.
EDIT FILTER DATA
Location:
[
Other2]
Name :
[ ]
EX :
[――――――――――] Edit
DM :
[――――――――――] Edit
BA :
[――――――――――] Edit
9
Press the MODE key to end the setting and to return to the
OPERATION screen.
To delete registered data
1
)
Move the cursor to “Location” using the
S
key or the
T
key and select data to delete using PREV or NEXT
key.
2
)
Press CLR key.