III Additional Functions
3 Restricting Rotation of Revolving Nosepiece
- 33 -
III
3
Restricting Rotation of Revolving Nosepiece
You can restrict the rotation of the revolving nosepiece with the OBJECTIVE key, so that the revolving nosepiece
will not turn from the set z-axis position and upward. The upper limit setting is effective for all the addresses of the
revolving nosepiece.
(1) Set Rotation Restriction
1
Press the MODE key, and the SETTING MENU screen will
appear on the LCD.
Select “More” using the
S
key or the
T
key and then
press the [
] key.
SETTING MENU
Objective
Condenser
Filter
Exciter
Barrier
Combination(Obj.>Condenser)
Combination(Filter>EX/BA)
More
2
Select “Rotation Stopper” using the
S
key or the
T
key
and then press the [
] key.
MORE
Rotation Stopper
Focus
Other Unit
Remote Control Pad
Shutter Key
Foot Switch
Communication Delimiter
3
Move the z-axis position to the position to restrict the
revolving nosepiece rotation (the position at which you
want the revolving nosepiece to stop turning).
4
Move the cursor to “ON” using the
W
key or the
X
key
and then press the [
] key.
ROTATION STOPPER
Current setting:OFF
ON OFF
5
A confirmation message appears.
Press the [
] key.
To cancel, move the cursor to “NO” using the
W
key or the
X
key and then press the [
] key.
ROTATION STOPPER
Current setting:ON
ON OFF
Rotation stopper ON ?
Yes No
6
When the setting is accepted, the “Complete!” message
appears and remains displayed for about one second.
ROTATION STOPPER
Current setting:ON
ON OFF
Complete!
7
Press the MODE key to end the setting and to return to the
OPERATION screen.
Summary of Contents for T-RCP