Checking and Adjusting the Plate Level
The axis of the plate level vial must be at right angles to the vertical axis of the
instrument.
To check and adjust the plate level:
1.
Set up the instrument on the tripod.
2.
Follow the leveling procedures described in
3.
Rotate the alidade 180°.
4.
Check whether the bubble is in the center of the vial.
5.
If the bubble is not in the center of the vial, adjust the plate level:
a.
Using the adjusting pin supplied, rotate
the plate level’s adjustment screw until
the bubble has moved half of the
distance back to the center.
b.
Using leveling screw A, move the
bubble into the center of the vial.
c.
Repeat from
.
Checking and Adjusting the Circular Level
Once you have checked and adjusted the plate
level, check the circular level.
If the bubble is not in the center of the level, use
the adjusting pin to rotate the three adjustment
screws until the bubble is centered.
C
B
A
2
Summary of Contents for NPL-322
Page 14: ...Contents xiv Total Station NPL 322 Instruction Manual ...
Page 38: ...2 Preparation 24 Total Station NPL 322 Instruction Manual ...
Page 68: ...3 Getting Started 54 Total Station NPL 322 Instruction Manual ...
Page 110: ...4 Applications 96 Total Station NPL 322 Instruction Manual ...
Page 154: ...5 Menu Key 140 Total Station NPL 322 Instruction Manual ...
Page 161: ...Total Station NPL 322 Instruction Manual 147 Checking and Adjustment 6 ...
Page 162: ...6 Checking and Adjustment 148 Total Station NPL 322 Instruction Manual ...
Page 172: ...8 System Diagrams 158 Total Station NPL 322 Instruction Manual ...
Page 188: ...9 Communications 174 Total Station NPL 322 Instruction Manual ...