Total Station NPL-322 Instruction Manual
Leveling
When you level the instrument, you make the vertical axis of the instrument exactly
vertical. To level the instrument, use the plate level.
To level the instrument:
1.
Loosen the upper plate clamp.
2.
Rotate the alidade until the plate level is
parallel with any two of the leveling screws
(B and C).
3.
Use leveling screws B and C to move the
bubble into the center of the level.
4.
Rotate the alidade approximately 90°.
5.
Use leveling screw A to move the bubble into
the center of the level.
6.
Repeat Step 1 through Step 5 to center the
bubble in both positions.
7.
Rotate the alidade 180°.
8.
If the bubble in the plate level remains
centered, the instrument is level. If the
bubble moves off center, adjust the plate
level. For detailed instructions, see
Checking and Adjusting the Plate Level,
.
C
B
A
1
C
B
A
2
Summary of Contents for NPL-322
Page 14: ...Contents xiv Total Station NPL 322 Instruction Manual ...
Page 38: ...2 Preparation 24 Total Station NPL 322 Instruction Manual ...
Page 68: ...3 Getting Started 54 Total Station NPL 322 Instruction Manual ...
Page 110: ...4 Applications 96 Total Station NPL 322 Instruction Manual ...
Page 154: ...5 Menu Key 140 Total Station NPL 322 Instruction Manual ...
Page 161: ...Total Station NPL 322 Instruction Manual 147 Checking and Adjustment 6 ...
Page 162: ...6 Checking and Adjustment 148 Total Station NPL 322 Instruction Manual ...
Page 172: ...8 System Diagrams 158 Total Station NPL 322 Instruction Manual ...
Page 188: ...9 Communications 174 Total Station NPL 322 Instruction Manual ...