Total Station NPL-322 Instruction Manual
Setting the Measurement Mode and Preparing the Target
The NPL-322 has two measurement modes: Prism mode (
Prism
) and Reflectorless
mode (
N-Prism
). These modes can be changed at any time by holding down the
[MSR1]
or
[MSR2]
key for one second. For more information, see
.
To set the measurement mode depending on the target you want to measure, see the
following table.
In some cases, you can measure another target that is not appropriate to the set
measurement mode.
Note –
The NPL-322 is Laser Class 1 in the measurement function, and Laser Class 2
in the Laser pointer function.
Don
’
t sight the Prism when the Laser Pointer is on.
Measurement with a prism
Do not use a prism with scratches, a dirty surface, or a chipped center. Prisms with
thin edges are recommended.
As the NPL-322 is extremely sensitive, multiple reflections on the prism surface can
sometimes cause a significant loss in accuracy.
To maintain the accuracy of your measurements:
Target
Target setting
Prism, reflector sheet
Prism (Prism mode)
Other (reflective materials)
N-Prism (Reflectorless mode)
thin edges
chipped center
thick edges
✓
✕
✕
Summary of Contents for NPL-322
Page 14: ...Contents xiv Total Station NPL 322 Instruction Manual ...
Page 38: ...2 Preparation 24 Total Station NPL 322 Instruction Manual ...
Page 68: ...3 Getting Started 54 Total Station NPL 322 Instruction Manual ...
Page 110: ...4 Applications 96 Total Station NPL 322 Instruction Manual ...
Page 154: ...5 Menu Key 140 Total Station NPL 322 Instruction Manual ...
Page 161: ...Total Station NPL 322 Instruction Manual 147 Checking and Adjustment 6 ...
Page 162: ...6 Checking and Adjustment 148 Total Station NPL 322 Instruction Manual ...
Page 172: ...8 System Diagrams 158 Total Station NPL 322 Instruction Manual ...
Page 188: ...9 Communications 174 Total Station NPL 322 Instruction Manual ...