49
VII. Specifications
Model name
Optical system
Light source
connection
Illumination
Field number
Illumination method
Illumination selection
method
Field diaphragm
Aperture diaphragm
Filters
Interface
Operating condition
Storage condition
LV-UEPI2A Motorized Universal Epi Illuminator 2A
CFI60 system (chromatic aberration free infinity optics system)
1 (rear)
Köhler illumination
25
Bright-field, dark-field, DIC*, simplified polarization*, sensitive color polarization*, epi-fl*
(* needs options)
Four port turret rotation in conjunction with a shutter to provide dazzling light
(noise terminator attached)
Turret drive method:
motor
Filter cube:
up to two filter cubes can be installed (filter cubes for the
bright-field and dark-field microscopy are fixed)
Adjustment:
manual
Variable range:
1.0 to 8.9 mm in diameter (bright-field microscopy)
Projected magnification: 3.0x (on the eyepiece image plane)
Full open diameter (for dark-field microscopy):
9.8 mm
Centering range:
2.4 mm in diameter
Adjustment:
motorized
Variable range:
1.2 to 8.0 mm in diameter (bright-field microscopy)
Projected magnification: 1.55x (on the objective pupil plane)
Full open diameter (for dark-field microscopy):
8.5 mm
Centering range:
2.4 mm in diameter
Built-in filter:
lemon skin filter,
UV filter (only for bright-field and dark-field microscopy)
ND filter:
two manual type sliders
Analyzer slot:
right side
Polarizer slot:
two slots on right side, lambda plate attachable
Excitation light balancer slot: left side
Special interface (for LV100DA or LV-ECON)
Temperature:
0°C to +40°C
Humidity:
85% relative humidity maximum (no condensation)
Altitude:
2000 m maximum
Pollution degree:
Degree 2
Installation category:
Category II
Electric shock protection class: Class III
Indoor use only
Temperature:
-20°C to +60°C
Humidity:
90% relative humidity maximum (no condensation)
LV-UEPI2A