31
IV. Operation
Modification from the registered condition (User offset)
When the illumination or the aperture diaphragm for the episcopic illumination is modified from the
condition registered in the set-up memory, the indicators on the operation panel blink indicating
that it is not in the registered condition any more. This modification is referred to as “User offset”
hereafter.
The condition of “User offset” is registered in the memory of this product: therefore, the post-
modified condition is restored when the registered combination of the microscopy and the objective
is selected, even after another microscopy or objective is selected or the power is turned off.
To clear “User offset”
To clear the user offset condition and return to the
registered conditions of the microscope system, hold down
the EPI button and the DIA button for two seconds or
longer.
When the registered conditions are restored, a short beep
sounds and the indicators on the operation panel turn on.
CAUTION
Even if the interlock control mode is changed, the user offset settings are kept.
Therefore, if the interlock control mode is changed with “LVSetup” under conditions
of user offset, the microscope system condition may differ from user intension. To
change the interlock control mode, perform the following if necessary and reset the
user offset settings beforehand.
For information about the interlock control mode, see Page 29. For information about
changing method of the interlock control mode, refer to the “LV Series Support Tools
software manual.”
To clear “User offset”
Hold down both buttons
for 2 seconds or more.