42
V. Troubleshooting
Problem
No interference color is
seen in the DIC
microscopy.
Uneven colors are seen
or low contrast image is
seen in the DIC
microscopy.
Countermeasure
Place the component into the optical
path.
(p.37)
Place the component into the optical
path.
(p.37)
Perform the setup work to specify the
settings properly.
(Refer to “LV Series Support
Tools software manual.”)
Use an objective that has the “LU Plan”
or “LU Plan Apo” mark.
Change the objective or the DIC prism
to a correct component.
Perform the setup work to specify the
settings properly.
(Refer to “LV Series Support
Tools software manual.”)
Cause
No analyzer or no polarizer is placed in
the optical path.
No DIC prism is placed in the optical
path.
The setup work has not performed. Or,
the setting is wrong.
A wrong objective is used.
The used objective and the DIC prism
do not match for the microscopy.
The setup work has not performed. Or,
the setting is wrong.