7.3 Performance tests
7-8
WF1966
Squarewave response
Connection:
Use coaxial cable to connect FUNCTION OUT to an oscilloscope (50
Ω
termination).
Setting:
Initialize, and then set
, the frequency to 1 MHz, the amplitude to 20 Vp-p/open,
and tr/tf to 7 ns.
Measurement:
Observe the waveform and measure the rise and fall times, overshoot and
undershoot.
Judgment:
The normal range is indicated in the table.
Item Normal
range
Rise and fall times
10 ns or less
Overshoot and undershoot
5 % or less
Duty factor
Connection:
Use coaxial cable to connect Function Out to a universal counter (50
Ω
terminated).
Settings:
Initialize the settings, then set amplitude to 20 Vp-p, and waveform and frequency
according to the following table.
Measurement:
Set the counter rise and fall period to the interval timer mode and measure the duty
(time). Use the average value, since jitter will cause dispersion in the measurement
values.
Determination:
The ranges indicated in the following table are normal.
Waveform Frequency Rated
range
(duty 50 % fixed)
1 MHz
490 to 510 ns
(duty 50 % fixed)
10 MHz
45.0 to 55.0 ns
(duty 50 % fixed)
50 MHz
6 to 14 ns
(duty variable)
100 kHz
4.90 to 5.10
µ
s
Summary of Contents for wf1966
Page 1: ...INSTRUCTION MANUAL NF Corporation WF1966 MULTIFUNCTION SYNTHESIZER...
Page 2: ......
Page 3: ...DA00008200 002 WF1966 2CH 50MHz SYNTHESIZER Instruction Manual...
Page 4: ......
Page 8: ......
Page 20: ......
Page 31: ...3 1 Panel description WF1966 3 3 Front panel...
Page 32: ...3 1 Panel description 3 4 WF1966 Rear panel...
Page 62: ......
Page 158: ......
Page 174: ......
Page 190: ......
Page 192: ......
Page 194: ......
Page 196: ......
Page 197: ......