WF1966
7-3
7.2 Operation checks
Preparatory checks
Check the following before inspection.
z
Power source voltage is within specified range.
z
Ambient temperature is within 5 to 35
°
C.
z
Ambient relative humidity is within 5 to 85 %RH (absolute humidity: 1 to 25 g/m
3
)
z
Condensation is absent.
Function checks
z
Power on
Confirm absence of error message at power on.
If an error message appears,
6. Troubleshooting.
If an abnormal indication appears at power on, switch power off and wait at least 5 seconds, then
again switch power on.
z
Main function checks
To avoid setting error, initialize the settings (
SYSTEM
→
PRESET).
Connect FUNCTION OUT to an oscilloscope with 50
Ω
coaxial cable to monitor the output.
Operate to change the following settings several times and confirm normal functions. Test both
keypad and MODIFY dial for settings such as frequency.
・
Frequency (
ENTRY
→
FREQ
)
・
Amplitude (
ENTRY
→
AMPTD
)
・
DC offset (
ENTRY
→
OFFSET
)
・
Waveform (
FUNCTION
→
,
,
Duty fixed 50 %
,
,
)
・
Duty (
FUNCTION
→
Variable duty
,
ENTRY
→
DUTY
)
・
Output on/off (
C H 2 O U T
O N /O F F
C H 1 O U T
O N /O F F
,
)
Summary of Contents for wf1966
Page 1: ...INSTRUCTION MANUAL NF Corporation WF1966 MULTIFUNCTION SYNTHESIZER...
Page 2: ......
Page 3: ...DA00008200 002 WF1966 2CH 50MHz SYNTHESIZER Instruction Manual...
Page 4: ......
Page 8: ......
Page 20: ......
Page 31: ...3 1 Panel description WF1966 3 3 Front panel...
Page 32: ...3 1 Panel description 3 4 WF1966 Rear panel...
Page 62: ......
Page 158: ......
Page 174: ......
Page 190: ......
Page 192: ......
Page 194: ......
Page 196: ......
Page 197: ......