![NF FRA5022 Instruction Manual Download Page 51](http://html1.mh-extra.com/html/nf/fra5022/fra5022_instruction-manual_1684892051.webp)
3.5
Basic Operations
FRA5022
3-21
■
Oscillator frequency
OSC frequency
Value range:
0.10mHz - 100.00kHz
Resolution: 5 digits, or 0.01mHz whichever
is the largest.
Description:
The oscillator frequency for spot measurements.
Upon establishing the frequency of spot measurement, the actual
output frequency changes immediately.
Even when the frequency is changed, the signal phase remains
consistent and the waveform is changed smoothly.
Initial values: 1kHz
■
Oscillator AC amplitude
OSC ac amplitude
Value range:
0.000 - 7.07Vrms
Resolution: 0.001 (
<
1), 0.01 (
≧
1) Vrms
0.000 - 10.00Vpeak Resolution: 0.001 (
<
1), 0.01 (
≧
1) Vpeak
Description:
The output AC amplitude of the oscillator. When the output terminal
is open this is the value that is used.
Changes in this value will not be reflected until the Output ON
AC/DC key is pressed.
Initial values: 10mVrms
■
Oscillator DC bias
OSC dc bias
Value range:
-
10.00V - +10.00V
Resolution: 0.01V
Description:
The output DC bias of the oscillator. When the output terminal is
open this is the value that is used.
Changes in this value will not be reflected until the Output ON
AC/DC key is pressed.
Initial values: 0V
Summary of Contents for FRA5022
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual...
Page 2: ......
Page 3: ...FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual DA00016981 005...
Page 26: ...2 2 Installation FRA5022 2 8 Figure 2 5 Removal of feet...
Page 30: ...FRA5022 2 12...
Page 92: ...5 5 Command Explanation FRA5022 5 14 Query example SRE Response example 128...
Page 119: ...FRA5022 5 41...
Page 131: ...FRA5022 7 8...
Page 141: ......
Page 143: ......
Page 144: ......