3.5
Basic Operations
FRA5022
3-17
3.5.4 System Menu Screen
Items that are common to
FRA5022
as a whole, or remote control items can be setup in the
System menu screen.
Figure 3-9
System menu screen
■
Measurement data recording
STORE data A-->B
Value range:
1) Execute
Description:
By pressing the 1 key, the most recent measurement data (data
memory A contents) is copied to data memory B.
■
Self calibration
CALIBRATION
Value range:
1) Execute, 9) Extend
Description:
By pressing the 1 key, a self calibration is executed.
Self calibration takes about 50 seconds. It is not possible to stop it
while it is being executed.
9) Extend is reserved for the function extensions. Currently is the
same function as 1) Execute.
Selected items'
settings or
setting range is
displayed.
Set values
Setup items:
select with the
Up/Down cursor
keys
.
Scrollbar:
when a setup item
cannot be fully
displayed on the
screen the position of
the item summary is
thus displayed.
Three sub-menus: switchable with the Left/Right cursor keys
△
▽
.
OSC slow on/off
= Quick only
0
)
Quick only 1)Slow enable
STORE data A-->B
CALIBRATION
SCAN limit
SCAN man/auto
OSC slow on/off
INTERFACE connect
INTERFACE address
Serial Number
9087651
2
GPIB
Quick only
Auto
3
M E N U < S Y S T E M > < S W E E P > < S P O T >
▲
▼
Summary of Contents for FRA5022
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual...
Page 2: ......
Page 3: ...FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual DA00016981 005...
Page 26: ...2 2 Installation FRA5022 2 8 Figure 2 5 Removal of feet...
Page 30: ...FRA5022 2 12...
Page 92: ...5 5 Command Explanation FRA5022 5 14 Query example SRE Response example 128...
Page 119: ...FRA5022 5 41...
Page 131: ...FRA5022 7 8...
Page 141: ......
Page 143: ......
Page 144: ......