![NF FRA5022 Instruction Manual Download Page 116](http://html1.mh-extra.com/html/nf/fra5022/fra5022_instruction-manual_1684892116.webp)
5.6 Status System
FRA5022
5-38
Table 5-5
Operation condition register contents
Bit
Weight
Contents (when standard status 1)
SPOT 12
4096
During spot measurement
Single: At each measurement, set to 1 when the measurement
starts, and cleared to 0 when the measurement ends.
Repeat: Remains 1 during a repeat measurement.
Not cleared to 0 at each measurement.
MSW 11
2048
In manual sweep measurement: running measurement at one
point
ASW 10
1024
Running automatic sweep measurement
HOLD 9
512
Hold state
SLOW 8
256
Changing output voltage SLOW
CAL
0
1
Running self calibration
Other
-
-
Always at 0 (unused)
■
Related commands / queries
STATus:OPERation:CONDition?
Queries the operation condition register contents.
The condition register content is not cleared to 0 even if queried.
Always indicates the status of
FRA5022
.
STATus:OPERation:NTRansition / STATus:OPERation:NTRansition?
Sets/queries the negative transition filter register.
STATus:OPERation:PTRansition / STATus:OPERation:PTRansition?
Sets/queries the positive transition filter register.
Set to 0 to clear these filter registers to 0.
Cannot be cleared by any other command.
Cleared to 0 upon powering on.
STATus:OPERation[:EVENt]?
Queries operation event register.
The event register is cleared to 0 if queried.
The event register can also be cleared with a
*
CLS command.
Cleared to 0 upon powering on.
STATus:OPERation:ENABle / STATus:OPERation:ENABle?
Sets/queries the operation event enable register.
Set to 0 to clear the enable register to 0.
Cannot be cleared by any other commnad.
Cleared to 0 upon powering on.
A parameter of each register setting message or response message takes a value that adds
up the weight of all bits set to 1.
Summary of Contents for FRA5022
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual...
Page 2: ......
Page 3: ...FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual DA00016981 005...
Page 26: ...2 2 Installation FRA5022 2 8 Figure 2 5 Removal of feet...
Page 30: ...FRA5022 2 12...
Page 92: ...5 5 Command Explanation FRA5022 5 14 Query example SRE Response example 128...
Page 119: ...FRA5022 5 41...
Page 131: ...FRA5022 7 8...
Page 141: ......
Page 143: ......
Page 144: ......