5.5 Command Explanation
FRA5022
5-29
[:SOURce]:VOLTage:OFFSet[:IMMediate] <dc_offset>
[:SOURce]:VOLTage:OFFSet[:IMMediate]?
Description:
Sets/queries the oscillator DC bias.
Parameters:
<
dc_offset
>
DC bias, format
<
NR2
>
,
range
-
10.00 to 10.00 [unit V]
Settings example:
VOLT:OFFS 10.00
Response:
<
dc_offset
>
Query example:
VOLT:OFFS?
Response example:
10.00
[:SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude] <amplitude> [VRMS|VPK]
[:SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
Description:
Sets/queries the oscillator AC amplitude.
The set value is interpreted as a value of the unit set in
VOLTage:UNIT.
When a suffix is used for this value (VRMS|VPK),
VOLTage:UNIT becomes irrelevant and the unit is interpreted
as being the suffix.
Parameters:
<
amplitude
>
AC amplitude :
<
NR2
>
0.000 to 7.07 [unit Vrms] |
0.000 to 10.00 [unit Vpk]
Settings example:
VOLT 7.07 Other examples) VOLT 10VPK, VOLT 1VRMS
Response:
<
amplitude
>
Query example:
VOLT?
Response example:
7.07
[:SOURce]:VOLTage:UNIT <amplitude_unit>
[:SOURce]:VOLTage:UNIT?
Description:
Sets/queries the oscillator AC amplitude display unit and
response unit.
Parameters:
<
amplitude_unit
>
AC amplitude unit, format
<
CRD
>
,
range VRMS | VPK
Settings example:
VOLT:UNIT VRMS
Response:
<
amplitude_unit
>
Query example:
VOLT:UNIT?
Response example:
VRMS
Summary of Contents for FRA5022
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual...
Page 2: ......
Page 3: ...FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual DA00016981 005...
Page 26: ...2 2 Installation FRA5022 2 8 Figure 2 5 Removal of feet...
Page 30: ...FRA5022 2 12...
Page 92: ...5 5 Command Explanation FRA5022 5 14 Query example SRE Response example 128...
Page 119: ...FRA5022 5 41...
Page 131: ...FRA5022 7 8...
Page 141: ......
Page 143: ......
Page 144: ......