3.5
Basic Operations
FRA5022
3-13
3.5
Basic Operations
3.5.1 Simplified Sweep Measurement
This section describes simplified operations when first using
FRA5022
.
It describes as frequency sweep and a measurement of a system's frequency response.
Proceed according to the following procedure.
1) Initialization: executed to simplify understanding of the following steps.
2) Setting the measurement conditions
3) Sweep measurement
4) Measured values reading
■
Initialization
・
Press the MENU/EXIT key to display the menu screen.
・
Use the Left/Right cursor keys
△
▽
to move to the system menu
<
SYSTEM
>
screen.
・
Use the Up/Down cursor keys to select INITIALIZE all.
・
Press the key 1 to initialize.
■
Setting the measurement conditions
・
Use the Left/Right cursor keys
△
▽
to move to the sweep setup
<
SWEEP
>
screen.
・
Set the values as below for example.
AC Amplitude
OSC ac amplitude
1.00Vrms Set the value with the ENTR key
DC bias
OSC dc bias
0.00V
Set the value with the ENTR key
Integration time
MEAS integ. time 0.10s
Set the value with the ENTR key
Integration cycle
MEAS integ. cycle 1c
Set the value with the ENTR key
Frequency max
SWEEP freq. max 100.00kHz Set the value with the k key
Frequency min
SWEEP freq. min 10.000Hz Set the value with the ENTR key
Automatic scaling DISP scale
Auto
After measurement, the vertical scale is
automatically selected.
For the possible settings range, check the comments in the lower side of the screen.
・
Press the ON AC/DC OSCILLATOR key to turn the oscillator's output on.
The oscillator's output voltage will not change by only modifying the setup value. It is
changed by this procedure.
■
Sweep measurement
・
Press the Measurement DOWN or UP key.
The screen changes to Sweep measurement screen and the measured gain and phase
are displayed as a graph (Bode diagram).
■
Measured values reading
・
Use the Left/Right cursor keys
△
▽
to move the graph's frequency data cursor
(vertical line), the measured value at the position of the cursor is displayed as a
numerical value.
Summary of Contents for FRA5022
Page 1: ...NF Corporation FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual...
Page 2: ......
Page 3: ...FREQUENCY RESPONSE ANALYZER FRA5022 Instruction Manual DA00016981 005...
Page 26: ...2 2 Installation FRA5022 2 8 Figure 2 5 Removal of feet...
Page 30: ...FRA5022 2 12...
Page 92: ...5 5 Command Explanation FRA5022 5 14 Query example SRE Response example 128...
Page 119: ...FRA5022 5 41...
Page 131: ...FRA5022 7 8...
Page 141: ......
Page 143: ......
Page 144: ......