Chapter 9 - Service
65
Safety inspection
More detailed inspection protocols are available from the manufacturer.
Content of the safety inspection (every 24 months):
1. visual examination
•
check for mechanical integrity
•
completeness of the product
•
labels (note the enclosed documents)
2. functional tests
•
turning on and off the switch
•
check of the different operation modes
•
simulation of electrodes slipping off of by changing the resistance
•
3. metrological test
•
measuring the output current of the DC-
STIMULATOR
•
check whether the accuracy limits are complied
4. safety control
•
according to the appropriate content of the EN 62353:2008
5. final acceptance test (FAT) as specified by the neuroConn GmbH
Please contact the manufacturer for detailed information to carry out the final
acceptance test.
The metrological control (every 12 months) includes the steps 1 - 3.