Chapter 5 - Software Reference Manual
44
Trigger input (optional)
To start stimulation remotely or via other devices, the DC-
STIMULATOR
can be used with
an external trigger. For technical details please see Manual Operation Basics: Trigger input.
Trigger input can be:
disabled (Standard)
Trigger input is disabled.
The stimulation can be started manually in STIMULATION menu (see Software Reference
Manual: STIMULATION). One stimulation only will be performed.
single
Trigger input is enabled for one stimulation.
After pressing "Y" in STIMULATION menu (see Software Reference Manual: STIMULATION),
the DC-
STIMULATOR
will wait for the trigger event to start the stimulation. One stimulation
will be performed after the trigger event.
repetitive
Trigger input is enabled for repetitive triggering.
After pressing "Y" in STIMULATION menu (see Software Reference Manual: STIMULATION),
the DC-
STIMULATOR
will wait for the trigger event to start the stimulation. One stimulation
will be performed after the trigger event.
After this stimulation the DC-
STIMULATOR
will wait for the next trigger event. This repetitive
procedure can be stopped by pressing the button 1 (X).
Figure 25: Main menu SYSTEM, submenu trigger input
To enable / disable the trigger input please follow these instructions (see Figure 25):
- Press button 2 or 4 to select main menu SYSTEM. Move the cursor to line 2 on the
display by pressing button 3.
- Switch to submenu trigger input by pressing button 2 or 4 several times. Move the cursor
to line 4 on the display by pressing button 3 (Figure 25).
Summary of Contents for DC-Stimulator
Page 1: ...Programmable Direct Current Stimulator DC STIMULATOR User manual...
Page 2: ......
Page 3: ......
Page 4: ......
Page 7: ...7 Table of contents 82 Electrode paste Ten20...
Page 26: ...Chapter 3 Getting Started 26 Figure 9 Turning off the DC STIMULATOR...
Page 80: ...Chapter 13 Notes 80 Notes 13...
Page 82: ...Chapter 14 Manuals Equipment 82 Electrode paste Ten20...