Chapter 4 - Operating Basics
34
External trigger source requirements
The external trigger source must use TTL levels:
low level
0-0.4 V
high level
2.4-5.25 V
min. current of trigger source
3 mA
The DC-
STIMULATOR
works:
- with
level detection
(not edge detection)
- with
"active high"
, i.e. stimulations are started by high levels
The trigger level must last for a minimum of
1 ms
.
Cleaning the device
Clean the DC-
STIMULATOR
using a cloth slightly dampened with water. For the display a
standard commercial TFT display cleaner should be used. Spray some cleaner on a soft
and clean cloth before wiping the display gently.
Never spray cleaner directly onto your DC-
STIMULATOR
and never let liquid run
into the device. Never use harsh or caustic chemicals to clean the DC-
STIMULATOR
. The manufacturer accepts no responsibility for any damage
caused by such use.
In general, disinfection of the system beyond normal cleaning is only necessary if people
from high-risk groups (e.g. drug users) are in contact with it. In this case, the careful use of a
common disinfection product is recommended to ensure user protection. Wipe the surface
of the DC-
STIMULATOR
and the electrode cables / supply cables with your disinfection
product (e.g. 90 % alcohol).
Summary of Contents for DC-Stimulator
Page 1: ...Programmable Direct Current Stimulator DC STIMULATOR User manual...
Page 2: ......
Page 3: ......
Page 4: ......
Page 7: ...7 Table of contents 82 Electrode paste Ten20...
Page 26: ...Chapter 3 Getting Started 26 Figure 9 Turning off the DC STIMULATOR...
Page 80: ...Chapter 13 Notes 80 Notes 13...
Page 82: ...Chapter 14 Manuals Equipment 82 Electrode paste Ten20...