Chapter 5 - Software Reference Manual
40
PARAMETER
This menu offers the selection of stimulation waveform modes and the associated
parameters like current, duration etc..
The chosen parameters can be saved into configurations (setting A - D) by pushing button 1
and 3 simultaneously. Stimulation parameters, which are not saved in a configuration, can
not be recalled on the next start-up.
To modify the stimulation waveform mode please follow these instructions:
- Press button 2 or 4 to select main menu PARAMETER.
- Move the cursor to line 2 on the display by pressing button 3.
- Switch to desired stimulation waveform mode by pressing button 2 or 4 several times.
To modify the parameters current, duration etc. please follow these instructions:
- After you have switched to the desired stimulation waveform mode move the cursor to
line 3 by pressing button 3.
- Press button 2 or 4 to select the desired parameter.
- Move the cursor to line 4 by pressing button 3.
- Pressing button 2 und 4 you can set the parameter. Press button 1 to confirm and save
the value.
tDCS
Using the stimulation waveform mode tDCS a transcranial direct current stimulation can be
applied.
A current of value
[current]
µ
A will last for
[duration]
seconds.
To avoid painful current steps the user can define controlled rising and falling of the DC by
adjusting
[fade in]
and
[fade out]
values (Figure 21). These fade in and fade out periods are
both additional to the stimulation's
[duration]
. The total stimulation time will therefore be
calculated as follows:
Summary of Contents for DC-Stimulator
Page 1: ...Programmable Direct Current Stimulator DC STIMULATOR User manual...
Page 2: ......
Page 3: ......
Page 4: ......
Page 7: ...7 Table of contents 82 Electrode paste Ten20...
Page 26: ...Chapter 3 Getting Started 26 Figure 9 Turning off the DC STIMULATOR...
Page 80: ...Chapter 13 Notes 80 Notes 13...
Page 82: ...Chapter 14 Manuals Equipment 82 Electrode paste Ten20...