PM Memory Error
This is a specific Power
Module based trip.
Check all cables and
connections. Reboot the
wheelchair.
Reprogram the Rnet system.
See section 5.3.
If the trip is still present after
the above checks have been
made, then the Power
Module may be defective.
See section 3.12.
Bad cable
This occurs when the control
system detects a fault in the
wiring
Check all cables and
connections for continuity. If
there is any visible damage
to cables, replace and reboot
the wheelchair. Disconnect
one cable from the system at
a time cycling the power
after each disconnection.
If the trip is still present after
the above checks have been
made, then the Power
Module may be defective.
See section 3.12.
Bad Settings
This occurs when the control
system detects incorrect or
invalid program settings.
Reprogram the Rnet system.
See section 5.3.
CxSM module error
9612
A problem occurred with the
cable between the CxSM and
the advanced seating
module (sensor board)
Reboot the wheelchair.
Check the cable is properly
inserted.
If the trip still continues
please contact the
manufacturer.
CxSM module error
9612
A problem occurred with
some actuators or cable.
Reboot the wheelchair.
Check the cables of actuators
are properly inserted.
If the trip still continues
please contact the
manufacturer.
Summary of Contents for EVO3
Page 1: ...SERVICE MANUAL Rev D 2014 10 11...
Page 2: ......
Page 4: ......
Page 36: ...www neatech it 36...
Page 69: ...www neatech it 69 5 2 Wiring diagram Figure 82...
Page 70: ...www neatech it 70 Figure 83...
Page 79: ...www neatech it 79 Figure 87...
Page 90: ...www neatech it 90 6 1 Dimensions...
Page 92: ......