Electronic Troubleshooting
Problems
Possible cause
Action
The wheelchair doesn’t start
Batteries discharged
Charge the batteries
Any Rnet cable not properly
inserted
Check all Rnet cables
The wheelchair cannot be
driven
Battery charger connected
Disconnect the battery
charger
Brake release lever not
correctly engaged
Set properly the brake
release lever (UP)
The wheelchair can only be
driven at reduced speed
The position of the seating
system is such that some
inhibition occurs
Set the wheelchair in the
DEFAULT POSITION, see
section 2.7.2
The wheelchair cannot be
charged
Main circuit breaker
switched off
Turn off the main circuit
breaker. If the problem
persists see section 5.6
5.6
Rnet system fault
When the control system is triggered and the wheelchair can’t be used, a diagnostic screen is
displayed in the display of the joystick.
First switch off the wheelchair and leave it off for some minutes. Then reboot the wheelchair. If
the fault still continues, contact an authorized service center.
Figure 92
Seeing Figure 35:
Identified Module indicates the module that had the problem
For example:
PM
Power Module
JSM
Joystick
ISM
Light Module
CxSM
Seating Module
The Trip Text is a brief description of the problem
The Trip Code is a four digit code that allow an authorized service center to identify the
problem
Summary of Contents for EVO3
Page 1: ...SERVICE MANUAL Rev D 2014 10 11...
Page 2: ......
Page 4: ......
Page 36: ...www neatech it 36...
Page 69: ...www neatech it 69 5 2 Wiring diagram Figure 82...
Page 70: ...www neatech it 70 Figure 83...
Page 79: ...www neatech it 79 Figure 87...
Page 90: ...www neatech it 90 6 1 Dimensions...
Page 92: ......