6-4
OLYMPIC STERILE-DRIER
Model 43/44 Instruction Manual
Repair Procedures
Replacing the Control Panel Printed Circuit Board
If the indicator lights don’t work, the blower doesn’t run, or the heater doesn’t
warm the air, the control panel printed circuit board (PCB) may need to
be replaced.
Required Items:
Control panel PCB (part no. 401123)
Screwdriver
Electrical grounding strap
To replace the control panel printed circuit board:
1
Press the power switch to the
OFF
(O) position, then unplug the power cord.
2
Open the door and remove the top four trays. Next, remove the acorn nut, bolts,
and washers that secure the tray brackets to the side access panel (see
Figure 6.1). Save all fasteners. Finally, remove the side access panel.
Figure 6.1
Control panel—exploded view
3
Remove the upper white door seal and loosen—but do not remove—the screws
just inside the Drier door behind the door stop (see Figure 4.1 on page 4-2).
4
Pull out and remove the front cover (see Figure 4.3 on page 4-3).
CAUTION
PCBs are electrostatic discharge (ESD) sensitive equipment. Use appropriate precautions
to prevent ESD.
NOTE
#
The stud (see Figure 6.1) is set into the frame and cannot be removed. Remove only the
acorn nut and washer.
Power switch
Front control
access panel
PCB
Stud
Side access panel
Kep nuts (x3)
Nuts and plastic
spacers (x6)
Acorn nut