2-6
OLYMPIC STERILE-DRIER
Model 43/44 Instruction Manual
Testing the Drier
Figure 2.7
Accessing the blower assembly
5
Separate the plenum from the casing (or HEPA filter), and then lift the plenum
and unscrew the heater from the socket. Ensure that the socket cap is tight, and
then screw the heater firmly back into place.
6
Replace the plenum; ensure that the plenum and filter are fully seated. Install
the plenum hold-down springs, and then reconnect the thermal cut-out wires.
7
Position the front cover, tighten the screws, and then push the white door seal
into place.
8
Plug in the power cord, and press the power switch to the
ON
(|) position.
9
Test for proper operation (see
).
Testing the Drier
Perform the test procedure to ensure the Drier is working properly after performing
assembly and service procedures.
To test the Drier:
1
Plug in the power cord, and press the power switch to the
ON
(|) position.
2
Set the heat control to midway between high and low, then set the timer to
ON
.
3
Ensure that the blowers are operating—warm air should blow from the screened
panel above the tube holder assembly inside the drying cabinet.
HEPA filter
FRONT
Acorn nuts and washers
Blower
Blower bracket
Blower filter plate
Plenum assembly
Note
: Pull together the two springs to
unlatch them, and then separate. Lift
the metal box up to access the heaters.