6-10
OLYMPIC STERILE-DRIER
Model 43/44 Instruction Manual
Repair Procedures
Replacing the Blower Control Relay
(Model 44 only)
If the timer is on, but the blower doesn’t operate, replace the blower control relay.
Required Items:
Blower control relay (part no. 200041)
Screwdriver
To replace the blower control relay:
1
Press the power switch to the
OFF
(O) position, then unplug the power cord.
2
Remove the upper white door seal and loosen—but no not remove—the screws
just inside the Drier door (behind the door stop) (see Figure 4.2 on page 4-3).
3
Pull out and remove the front cover (see Figure 4.3 on page 4-3).
4
Unplug the blower control relay from the socket (see drawing 700203).
5
Remove the old relay, then place the new relay into position.
6
Position the front cover, tighten the screws, then push the door seal into place.
7
Plug in the power cord, then press the power switch to the
ON
(|) position.
8
Test the device for proper operation (see page 2-4).
WARNING
Electrical shock hazard when the front panel is removed.