6-16
OLYMPIC STERILE-DRIER
Model 43/44 Instruction Manual
Repair Procedures
3
Pull out and remove the front cover (see Figure 4.3 on page 4-3).
4
Unhook, remove, and set aside the plenum hold-down springs, then disconnect
the thermal cut-out wires (see Figure 6.4 on page 6-12). Next, separate the
plenum from the filter.
5
Locate and remove the screws and kep nuts that secure the failed thermal
cut-out to the plenum assembly. Save the screws and kep nuts.
6
Place and secure the new thermal cut-out using the screws and kep nuts saved in
the previous step.
7
Fully seat the plenum and filter. Install the plenum hold-down springs, then
reconnect the thermal cut-out wires.
WARNING
Electrical shock hazard when the front panel is removed.
NOTE
#
In Model 44, the left and right plenum assemblies are mirror images of one another. For
example, the thermal cut-out on the right plenum assembly is located on the back, while
that on the left is visible from the front.
CAUTION
Avoid damaging the HEPA filter. Holes, tears, and rips adversely affect filter function.