OLYMPIC STERILE-DRIER
Model 43/44 Instruction Manual
6-11
Repair Procedures
Replacing the Blower/Blower Motor
If the blower or its motor fail, it should be replaced.
Required Items:
Blower, 50Hz (part no. 401163) or 60Hz (part no. 200036)
or
Blower motor (part no. P91278)
Allen or Torx driver
Screwdriver
Loctite
®
242 locking compound or equivalent
To replace the blower:
1
Press the power switch to the
OFF
(O) position, then unplug the power cord.
2
Remove the upper white door seal, then loosen—but don’t remove—the screws
just inside the Drier door, behind the door stop (see Figure 4.2 on page 4-3).
3
Pull out and remove the front cover (see Figure 4.3 on page 4-3).
4
From inside the drying cabinet, remove and set aside the four acorn nuts and
four lock washers that secure the blower bracket to the blower filter plate (see
NOTE
#
For 60Hz devices only, the blower motor (rather than the entire blower assembly) may be
replaced, follow step 8a through step 8h.
WARNING
Electrical shock hazard when the front panel is removed.