OLYMPIC STERILE-DRIER
Model 43/44 Instruction Manual
Periodic Maintenance
Periodic Maintenance
Changing the Prefilter
To keep the drier in optimal operating condition, change the prefilter every two to
three months.
Required Items:
Screwdriver
Prefilter (set of 6) (part no. 54308)
To change the prefilter:
1
Loosen the bottom screws of the louver (see Figure 4.1).
Figure 4.1
Accessing the prefilter
2
Lift the bottom edge of the louver, then remove the prefilter.
3
Install the new prefilter with the air flow direction arrow pointing toward the
plenum. The airflow direction arrow is on the side of the prefilter.
4
Close the louver, then tighten the bottom screws.
Semi-Annual Maintenance
Lubricating the Blowers
To keep the Drier in optimal operating condition, lubricate the blowers every
six months.
Required Items:
SAE 20 oil
Screwdriver
To lubricate the blowers:
1
Press the power switch to the
OFF
(O) position, then unplug the power cord.
2
Remove the upper white door seal, then loosen—but don’t remove—the screws
just inside the Drier door, behind the door stop (see Figure 4.2).