National Instruments Corporation
21
NI 9219 Calibration Procedure
Figure 13.
Half-Bridge Accuracy Verification Connections to the NI 9219 (Configuration 3)
Figure 14.
Half-Bridge Accuracy Verification Connections to the NI 9219 (Configuration 4)
9.
Configure DMM1 for a voltage measurement in the 10 V range.
10. Enable Auto Zero on DMM1.
11. Configure DMM2 for a voltage measurement in the 10 V range.
12. Enable Auto Zero on DMM2.
HI
V
LO
DMM2
NI 9219
HI
V
LO
DMM1
V+
V–
Source
EX+ (3)
HI (4)
EX– (5)
Ch
x
1 k
Ω
± 1%
1 k
Ω
± 1%
5 k
Ω
± 1%
5 k
Ω
± 1%
HI
V
LO
DMM2
NI 9219
HI
V
LO
DMM1
V+
V–
Source
EX+ (3)
HI (4)
EX– (5)
Ch
x
1 k
Ω
± 1%
1 k
Ω
± 1%
5 k
Ω
± 1%
5 k
Ω
± 1%