NI 9219 Calibration Procedure
34
ni.com
5.
Obtain an array of the recommended calibration resistances for your device using the NI 9219
instance of the DAQmx Get C Series Calibration Adjustment Points VI.
6.
Place the NI 9219 channel in voltage mode using the NI 9219 instance of the DAQmx Setup
C Series Calibration VI.
7.
Configure the DMM for a 4-wire resistance measurement.
8.
Enable Auto Zero on the DMM.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxGet9219CalAdjustPoints
with
the following parameters:
calHandle
:
calHandle
adjustmentPoints
: An array of reference values
for the DAQmx Adjust Calibration VI.
bufferSize
: The size of the
adjustmentPoints
array in elements.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxSetup9219Cal
with the following
parameters:
rangeMin
:
–1
rangeMax
:
1
calHandle
:
calHandle
channelNames
:
cDAQ1Mod8/aix
*
measType
:
DAQmx_Val_Voltage
bridgeConfig
:
DAQmx_Val_NoBridge
*
x
refers to the channel number