National Instruments Corporation
35
NI 9219 Calibration Procedure
9.
Acquire a resistance reading with the DMM.
10. Disconnect the DMM from the resistors and NI 9219.
11. Set up the signal acquisition using the NI 9219 instance of the DAQmx Setup C Series
Calibration VI.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxSetup9219Cal
with the following
parameters:
rangeMin
:
0
rangeMax
:
1050, 10500
calHandle
:
calHandle
channelName
:
cDAQ1Mod8/a
i
x
*
measType
:
DAQmx_Val_Resistance
bridgeConfig
:
DAQmx_Val_NoBridge
*
x
refers to the channel number