NI 9219 Calibration Procedure
44
ni.com
15. Acquire voltage measurements with both DMMs. Perform the following calculation using the
two DMM measurements.
16. Acquire a bridge reading with the NI 9219 using the NI 9219 instance of the DAQmx Adjust
C Series Calibration VI.
17. Repeat steps 10 through 16 for every value in the array.
18. Compute and save the adjustment to the EEPROM using the DAQmx Close External
Calibration VI. This VI also saves the date, time, and temperature of the adjustment to the
EEPROM.
19. Disconnect the DMMs and the resistors from the device.
20. Repeat steps 1 through 19 for each channel on the NI 9219.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxAdjust9219Cal
with the following
parameters:
calHandle
:
calHandle
channelName
:
cDAQMod8/ai
x
*
value
:
Ratio
ref
.
*
x
refers to the channel number
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxCloseExtCal
with the following
parameters:
calHandle
:
calHandle
action
:
DAQmx_Val_Action_Commit
Ratio
ref
V
DMM
2
V
DMM
1
----------------
=