NI 9219 Calibration Procedure
28
ni.com
10. Acquire a voltage reading with the NI 9219 using the NI 9219 instance of the DAQmx Adjust
C Series Calibration VI.
11. Repeat steps 6 through 10 for every value in the array.
12. Compute and save the adjustment to the EEPROM using the DAQmx Close External Calibration
VI. This VI also saves the date, time, and temperature of the adjustment to the EEPROM.
13. Repeat steps 3 through 12 for each voltage range on the NI 9219 channel.
14. Repeat steps 3 through 13 for each channel on the NI 9219.
15. Disconnect the source and the DMM from the device.
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxAdjust9219Cal
with the following
parameters:
calHandle
:
calHandle
channelName
:
cDAQ1Mod8/aix
*
value
: The DMM reading.
*
x
refers to the channel number
LabVIEW Block Diagram
NI-DAQmx Function Call
Call
DAQmxCloseExtCal
with the following
parameters:
calHandle
:
calHandle
action
:
DAQmx_Val_Action_Commit